Quality, Medical Devices & Electrical
ISO 13485, ISO 14971, IEC 62366, IEC 62304, ISO 26782, ISO 23747, IEC 60601-1, IEC 60601-2, ISO 10993-1
FDA
510(k) clearance
Associations & Institutes
ATS/ ERS 2005, NIOSH, OSHA
Available languages
English, French, German, Spanish, Italian, Dutch, Swedish, Danish, Portuguese
Printing options
Direct to printer or with EasyOne Connect software
Data management
EasyOne Connect
Export
HL7, XML, GDT, with software
Data links
USB, Bluetooth
Test storage
Up to 10‘000 tests
Age range
Spirometry > 4 years
Dimensions
87 x 155 x 36 mm (H x W x D), 356 g, 3.4 x 6.1 x 1.4‘‘ (H x W x D), 13 oz
Device classification
Type BF applied part
Operating conditions
Temp 0 - 40 °C/ 32 - 104 °F, Rel. Humidity 5 - 90%, Atmosph. Pressure 700 - 1060 hPa
Power supply
5 VDC, Standby 0.3W
Rechargeable battery
Exchangeable, 3.6 VDC
FVC
BEV, EOTV, FEF10, FEF25, FEF 2575, FEF2575/FVC, FEF40, FEF50, FEF50/FVC, FEF60, FEF75, FEF80, FET, FET2575, FEV.25, FEV.5, FEV.5/FVC, FEV.75, FEV.75/FVC, FEV1, FEV1/FEV6, FEV1/FVC, FEV3, FEV3/FVC, FEV6, FVC, MEF20, MEF25, MEF40, MEF50, MEF60, MEF75, MEF90, MMEF, PEF, PEFT, t0
FVL
BEV, EOTV, FEF10, FEF25, FEF 2575, FEF2575/FVC, FEF40, FEF50, FEF50/FVC, FEF60, FEF75, FEF80, FET, FET2575, FEV.25, FEV.5, FEV.5/FVC, FEV.75, FEV.75/FVC, FEV1, FEV1/FEV6, FEV1/FIV1, FEV1/FVC, FEV3, FEV3/FVC, FEV6, FIF25, FIF50, FIF50/FEF50, FIF75, FIV.25, FIV.5, FIV1, FVC, MEF20, MEF25, MEF40, MEF50, MEF60, MEF75, MEF90, MIF25, IF50, MIF75, MMEF, PEF, PEFT, t0
SVC
ERV, IC, IRV, Rf, VC, VCex, VCin, VCmax, VT
MVV
MVV, MVV6, MVVtime, VT
GLI
Stanojevic 2009, Quanjer 2012
North America
NHANES III (Hankinson) 1999, Knudson 1983, Knudson 1976, Crapo 1981, Morris 1971 & 1976, Hsu 1979, Dockery (Harvard) 1993, Polgar 1971, Gutierrez (Canada) 2004, Eigen 2001
Latin America
Pereira 1992, Perreira 2006 & 2008, Pérez-Padilla (PLATINO) 2006, Pérez-Padilla (Mexico) 2001, Pérez-Padilla (Mexico, Pediatrics) 2003, Chile 2010, Chile (Pediatrics) 1997
Europe
ERS (ECCS, EGKS, Quanjer) 1993, Zapletal 1977, Zapletal 2003, Rosenthal 1993, Austria 1988, Austria 1994, Sapaldia (Switzerland) 1996, Roca (Spain, SEPAR) 1982, Garcia-Rio (SEPAR) 2013, Vilozni 2005, Falaschetti 2004, Klement (Russia) 1986
Europe Scandinavia
Hedenström 1985 & 1986, Gulsvik (Norway) 1985, Berglund Birath (Sweden) 1963, Langhammer (Norway) 2001, Finnish 1982 (1998), Nystad 2002
Australia
Hibbert 1989, Gore Crockett 1995
Asia
Chhabra (India) 2014, Dejsomritrutai (Thailand) 2000, Indonesia 1992, IP (China, HongKong) 2000 & 2006, JRS 2001 & 2014
Africa
Ethiopia 1985
Type
Ultrasonic transit time
Flow Range
± 16 l/s
Flow Resolution
4 ml/s
Flow Accuracy (except PEF)
± 2% or 0.020 l/s
PEF Accuracy
± 5% or 0.200 l/s
MVV Accuracy
± 5% or 5 l/min
Volume Range
± 12 l
Volume Resolution
1  ml
Volume Accuracy
± 2% or 0.050 l
Resistance
0.3 cm H20/l/s at 16 l/s
Part Numbers
Description
2500-2A
EasyOne Air US Type A package
Part Numbers
Description
5050-50
EasyOne FlowTube, individually wrapped, box of 50 pcs
5050-200
EasyOne FlowTube, individually wrapped, box of 200 pcs
2030-2
ndd Calibration syringe 3L with EasyOne FlowTube Cal Check Adapter
5030-2
EasyOne FlowTube Adapter